Comparison of macular thickness in diabetic macular edema using spectral-domain optical coherence tomography and time-domain optical coherence tomography

J Ophthalmol. 2012:2012:959721. doi: 10.1155/2012/959721. Epub 2012 Jan 18.

Abstract

Purpose. To compare the macular thicknesses in diabetic macular edema (DME) measured with spectral-domain optical coherence tomography (SD-OCT) and time-domain (TD) OCT. Patients and Methods. The average macular thicknesses of 50 eyes of 29 patients with DME were measured using SD-OCT and TD-OCT. Results. The mean macular thicknesses measured with TD-OCT and SD-OCT were 401.5 ± 117.8 μm (mean ± SD) and 446.2 ± 123.5 μm, respectively. The macular thicknesses measured with the two devices were well correlated (Pearson's product moment correlation, r = 0.977, P < 0.001). A significant correlation was found between the best-corrected visual acuity and the retinal thickness measured by TD-OCT and SD-OCT (Pearson's product moment correlation, TD-OCT, r = 0.34; P < 0.05; SD-OCT, r = 0.32; P < 0.05). Discussion. The mean macular thickness measured with SD-OCT was about 45 μm thicker than that measured with TD-OCT. Attention should be paid when comparing data obtained using different OCT machines.