PREVAIL TRANSAPICAL: multicentre trial of transcatheter aortic valve implantation using the newly designed bioprosthesis (SAPIEN-XT) and delivery system (ASCENDRA-II)

Eur J Cardiothorac Surg. 2012 Aug;42(2):278-83; discussion 283. doi: 10.1093/ejcts/ezr325. Epub 2012 Jan 30.

Abstract

Objectives: Transapical (TA) aortic valve implantation (AVI) has evolved as an alternative procedure for high-risk patients. We evaluated the second-generation SAPIEN XT™ prosthesis in a prospective multicentre clinical trial.

Methods: A total of 150 patients (age: 81.6 ± 5.8 years; 40.7% female) were included. Prosthetic valves (diameter: 23 mm (n = 36), 26 mm (n = 57) and 29 mm (n = 57)) were implanted. The ASCENDRA-II™ modified delivery system was used in the smaller sizes. Mean logistic EuroSCORE was 24.3 ± 7.0%, and mean STS score 7.5 ± 4.4%. All patients gave written informed consent.

Results: Off-pump AVI was performed using femoral arterial and venous access wires as a safety net. All but two patients received TA-AVI, as planned. The 29-mm valve showed similar function as the values of two other diameters did. Three patients (2%) required temporary cardiopulmonary bypass support. Postoperative complications included renal failure requiring long-term dialysis in four, bleeding requiring rethoracotomy in four, respiratory complication requiring reintubation in eight and sepsis in four patients, respectively. Thirty-day mortality was 13 (8.7%) for the total cohort and 2/57 (3.5%) for patients receiving the 29-mm valve, respectively. Echocardiography at discharge showed none or trivial aortic incompetence (AI) in 71% and mild-AI in 22% of the patients. Post-implantation AI was predominantly paravalvular and ≥ 2+ in 7% of patients. One patient required reoperation for AI within 30 days.

Conclusions: The PREVAIL TA multicentre trial demonstrates good functionality and good outcomes for TA-AVI, using the SAPIEN XT™ prosthesis and its second-generation ASCENDRA-II™ delivery system, as well successful introduction of the 29-mm SAPIEN XT™ valve for the benefit of high-risk elderly patients.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged, 80 and over
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis*
  • Cardiac Catheterization / methods*
  • Female
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Prospective Studies
  • Prosthesis Design
  • Treatment Outcome