Gas pre-warming for improving performances of heated humidifiers in neonatal ventilation

Annu Int Conf IEEE Eng Med Biol Soc. 2011:2011:1205-8. doi: 10.1109/IEMBS.2011.6090283.

Abstract

Adequate temperature and humidification of gas delivered must be performed during long term neonatal ventilation to avoid potential adverse health effects. Literature shows that performances of heated humidifiers are, at least in some cases, quite poor. In this study, a novel approach to gas conditioning, consisting of gas warming upstream the humidification chamber, is presented. Gas pre-warming, in combination with a control strategy based on a mathematical model taking into account a number of parameters, allows to significantly improve the heated humidifier performances. The theoretical model has been validated and experimental trials have been carried out in the whole volumetric flow-rate (Q) range of neonatal ventilation (lower than 10 L · min(-1)). Experimental results (temperature values ranging from 36 °C to 38 °C and relative humidity values from 90 % to 98 % in the whole range of Q) show values very close to the ideal thermo-hygrometric conditions. The proposed solution allows to avoid vapor condensation at low flow rates and decrease of relative humidity at high flow rates.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Female
  • Gases / administration & dosage*
  • Heating / instrumentation*
  • Humans
  • Infant, Newborn
  • Male
  • Nebulizers and Vaporizers*
  • Respiration, Artificial / instrumentation*

Substances

  • Gases