Determination of three alkylphenol isomers in various water samples using a new HPLC method based on a duet stationary phase

Environ Monit Assess. 2012 Oct;184(10):6061-70. doi: 10.1007/s10661-011-2403-1. Epub 2011 Nov 10.

Abstract

The level of three alkylphenols (APs) 4-nonylphenol, 4-octylphenol and 4-tert-octylphenol was monitored in various water samples using a new developed and validated HPLC method. The HPLC method employed a column with a mixed-mode stationary phase (C18/SCX) using a mobile phase of water to methanol = 15:85 (v/v) delivered at a flow rate of 1 mL/min at room temperature. Both diode array, DAD and fluorescence, FLD detectors were employed. The method is linear when APs concentration ranged from 0.025 to 0.5 μg/mL with a DAD detection at 279 nm and between 0.0008 and 0.1 μg/mL when the detection was made by FLD (excitation at 220 nm, emission at 315 nm). The limit of detection and limit of quantitation for alkylphenols were found to be 5 and 15 ng/mL, respectively (detection by DAD). The method was employed with good results for the determination of APs in the presence of polycyclic aromatic hydrocarbons in various water samples.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid
  • Environmental Monitoring
  • Phenols / analysis*
  • Phenols / chemistry
  • Water Pollutants, Chemical / analysis*
  • Water Pollutants, Chemical / chemistry

Substances

  • Phenols
  • Water Pollutants, Chemical
  • 4-octylphenol
  • 4-nonylphenol
  • 4-tert-octylphenol