In vivo study of flow-rate accuracy of the MedStream Programmable Infusion System

Neuromodulation. 2011 May-Jun;14(3):235-41; discussion 241. doi: 10.1111/j.1525-1403.2011.00353.x. Epub 2011 Apr 15.

Abstract

Objective: Flow-rate accuracy and precision are important parameters to optimizing the efficacy of programmable intrathecal (IT) infusion pump delivery systems. Current programmable IT pumps are accurate within ±14.5% of their programmed infusion rate when assessed under ideal environmental conditions and specific flow-rate settings in vitro. We assessed the flow-rate accuracy of a novel programmable pump system across its entire flow-rate range under typical conditions in sheep (in vivo) and nominal conditions in vitro.

Materials and methods: The flow-rate accuracy of the MedStream Programmable Pump was assessed in both the in vivo and in vitro settings. In vivo flow-rate accuracy was assessed in 16 sheep at various flow-rates (producing 90 flow intervals) more than 90 ± 3 days. Pumps were then explanted, re-sterilized and in vitro flow-rate accuracy was assessed at 37°C and 1013 mBar (80 flow intervals).

Results: In vivo (sheep body temperatures 38.1°C-39.8°C), mean ± SD flow-rate error was 9.32% ± 9.27% and mean ± SD leak-rate was 0.028 ± 0.08 mL/day. Following explantation, mean in vitro flow-rate error and leak-rate were -1.05% ± 2.55% and 0.003 ± 0.004 mL/day (37°C, 1013 mBar), respectively.

Conclusions: The MedStream Programmable Pump demonstrated high flow-rate accuracy when tested in vivo and in vitro at normal body temperature and environmental pressure as well as when tested in vivo at variable sheep body temperature. The flow-rate accuracy of the MedStream Programmable Pump across its flow-rate range, compares favorably to the accuracy of current clinically utilized programmable IT infusion pumps reported at specific flow-rate settings and conditions.

Publication types

  • Evaluation Study

MeSH terms

  • Analgesics / administration & dosage
  • Analgesics / therapeutic use
  • Animals
  • Body Temperature
  • Chronic Pain / drug therapy
  • Humans
  • Infusion Pumps, Implantable / standards*
  • Infusions, Spinal*
  • Muscle Relaxants, Central / administration & dosage
  • Muscle Relaxants, Central / therapeutic use
  • Muscle Spasticity / drug therapy
  • Reproducibility of Results
  • Sheep

Substances

  • Analgesics
  • Muscle Relaxants, Central