Stability-Indicating RP-HPLC Method for Estimation of Miglitol in Bulk and Tablets

Indian J Pharm Sci. 2010 Nov;72(6):781-4. doi: 10.4103/0250-474X.84594.

Abstract

A selective and sensitive, stability-indicating reverse phase high performance liquid chromatography method has been first developed and validated for the estimation of miglitol in bulk and tablet dosages form. Samples were separated on a prepacked, Inertsil amino C(18) column (150×4.6 mm i.d.) using a mobile phase comprised of acetonitrile and monobasic sodium phosphate pH 7.5 (80:20, v/v) delivered at 1.5 ml/min flow rate. Detection was performed on a SPD-20A prominence UV/Vis detector at 220 nm. The retention time for miglitol was 13.93±0.0367. The method was validated in terms of linearity, precision, accuracy, ruggedness, and specificity, limit of detection and limit of quantification. The linearity (r(2)) and percentage recoveries of miglitol were 0.9986 and 99.85%. This method is suitable for routine estimation of miglitol in bulk and tablet dosages form.

Keywords: Miglitol; RP-HPLC; stability indicating and method validation.