Half-time myocardial perfusion SPECT imaging with attenuation and Monte Carlo-based scatter correction

Nucl Med Commun. 2011 Nov;32(11):1040-5. doi: 10.1097/MNM.0b013e328349c765.

Abstract

Purpose: To test the potential of a new reconstruction algorithm with Monte Carlo-based scatter correction in half-time myocardial perfusion single-photon emission computed tomography (SPECT).

Materials and methods: The mathematical four-dimensional NURBS-based Cardiac-Torso phantom and the SIMIND Monte Carlo simulation package were used to simulate full-time and half-time SPECT projection data. The data were reconstructed using the standard ordered subset expectation maximization-based algorithm and the new Monte Carlo-based algorithm. Defect contrast, myocardium versus ventricle contrast and resolution were calculated. In addition to the simulation studies, full-time and half-time SPECT projection data of 30 patients were reconstructed with the standard and the new method. The patient data were qualitatively evaluated by four nuclear medicine experts on a scale from 1 (poor quality) to 5 (high quality).

Results: The new reconstruction method with half-time data produced higher contrast and better resolution in the simulations and also achieved higher qualitative scores in the patient study than the standard reconstruction with full-time data.

Conclusion: Half-time myocardial perfusion imaging using the new reconstruction algorithm with Monte Carlo-based scatter correction produced images with superior quality when compared with full-time imaging with standard reconstruction.

MeSH terms

  • Female
  • Heart / diagnostic imaging*
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods
  • Male
  • Monte Carlo Method*
  • Myocardial Perfusion Imaging / methods*
  • Myocardium
  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • Perfusion
  • Phantoms, Imaging
  • Radiography
  • Reproducibility of Results
  • Scattering, Radiation
  • Sensitivity and Specificity
  • Tomography, Emission-Computed, Single-Photon / methods*

Substances

  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • technetium tc-99m tetrofosmin