Stent graft modification with mini-cuff reinforced fenestrations for urgent repair of thoracoabdominal aortic aneurysms

J Vasc Surg. 2011 Nov;54(5):1522-6. doi: 10.1016/j.jvs.2011.06.023. Epub 2011 Sep 16.

Abstract

Fenestrated endografts require 6 to 8 weeks for device customization, and off-the-shelf devices are not yet available and may not be of easy access for urgent repair of complex aneurysms. We describe a technique of stent graft modification in a high-risk male patient with two prior open aortic repairs, end ileostomy, and a rapidly enlarging 10-cm supra-graft type IV thoracoabdominal aortic aneurysm. A Z-stent thoracic stent graft was modified on-site using mini-cuff reinforced fenestrations to incorporate the visceral arteries and improve overlap at the side branch attachment sites. After successful repair, the patient was discharged at 4 days without complications and with patent branched stent grafts without endoleak. On-site modifications of endografts may allow urgent endovascular treatment of complex aortic aneurysms in high-risk patients who are not good candidates for open repair or who do not have access to manufactured fenestrated devices.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Aortic Aneurysm, Thoracic / diagnostic imaging
  • Aortic Aneurysm, Thoracic / surgery*
  • Aortography / methods
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Endovascular Procedures / instrumentation*
  • Humans
  • Male
  • Prosthesis Design
  • Stents*
  • Tomography, X-Ray Computed
  • Treatment Outcome