Cation exchange HPLC analysis of desmosines in elastin hydrolysates

Anal Bioanal Chem. 2011 Nov;401(8):2473-9. doi: 10.1007/s00216-011-5346-z. Epub 2011 Sep 2.

Abstract

Desmosine crosslinks are responsible for the elastic properties of connective tissues in lungs and cardiovascular system and are often compromised in disease states. We developed a new, fast, and simple cation exchange HPLC assay for the analysis of desmosine and isodesmosine in animal elastin. The method was validated by determining linearity, accuracy, precision, and desmosines stability and was applied to measure levels of desmosines in porcine and murine organs. The detection and quantification limits were 2 and 4 pmol, respectively. The run-time was 8 min. Our cation exchange column does not separate desmosine and isodesmosine, but their level can be quantified from absorbance at different wavelengths. Using this assay, we found that desmosines levels were significantly lower in elastin isolated from various organs of immunodeficient severe combined immunodeficiency mice compared with wild-type animals. We also found that desmosines levels were lower in lung elastin isolated from hyperhomocysteinemic Pcft(-/-) mice deficient in intestinal folate transport compared with wild-type Pcft(+/+) animals.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Chromatography, High Pressure Liquid / economics
  • Chromatography, High Pressure Liquid / methods*
  • Chromatography, Ion Exchange / economics
  • Chromatography, Ion Exchange / methods*
  • Desmosine / analysis*
  • Elastin / chemistry*
  • Hyperhomocysteinemia / metabolism
  • Isodesmosine / analysis*
  • Limit of Detection
  • Lung / chemistry
  • Male
  • Mice
  • Mice, Inbred C57BL
  • Mice, SCID
  • Severe Combined Immunodeficiency / metabolism
  • Swine
  • Time Factors

Substances

  • Desmosine
  • Elastin
  • Isodesmosine