Image quality and patient dose in computed tomography examinations in Greece

Radiat Prot Dosimetry. 2011 Sep;147(1-2):129-32. doi: 10.1093/rpd/ncr282. Epub 2011 Jul 9.

Abstract

The purpose of this study is to evaluate image quality of various computed tomography (CT) scanners installed in Greece, as well as to investigate patient doses from common CT examinations. An image quality survey was performed in 44 CT scanners countrywide. The imaging performance of the systems was evaluated by measurements of certain parameters, such as image noise, spatial uniformity, high- and low-contrast resolution and slice thickness accuracy. Moreover, preliminary results of patient dose survey are presented. Concerning image quality, 80 % of the scanners were found to be in compliance with the national legislation and relative international guidelines for all the examined parameters. Weighted CT dose index and dose-length product values for chest and abdomen routine examinations were generally below the dose reference levels (DRLs) suggested by the European Commission. However, some scanners were found to deliver significantly higher doses than the suggested DRL for head routine examinations. Finally, differences in the performance among scanners of the same type and similar age were observed, pointing out the importance of frequent calibration, routine quality control and proper maintenance.