[Evaluation of stereometric parameters of optic disc and nerve fiber layer using HRT III. Report 2: factors influencing reproducibility]

Vestn Oftalmol. 2011 Mar-Apr;127(2):43-6.
[Article in Russian]

Abstract

29 patients with initial primary open angle glaucoma were examined using Heidelberg retinal tomograph III (HRT III), factors influencing measurements reproducibility and variability were studied. The following factors were found to cause outliers in stereometric parameters of the optic nerve head and retinal nerve fiber layer: standard reference height showing large difference of measurements or very high values, significant difference of scan depth, small optic disc, astigmatism, old age, interobserver variability of disc area (contour line position). Considering these factors one can determine indications for control examination thus improving method accuracy.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Aged
  • Diagnostic Techniques, Ophthalmological
  • Follow-Up Studies
  • Glaucoma, Open-Angle / pathology*
  • Humans
  • Nerve Fibers / pathology*
  • Ophthalmoscopy / methods*
  • Optic Disk / pathology*
  • ROC Curve
  • Severity of Illness Index
  • Tomography / methods*
  • Tomography, Optical Coherence / methods