Background: The investigation and effective management of phaeochromocytoma involves biochemical measurement of either conjugated total urine or plasma free metadrenalines. Current analytical methods include enzyme-linked immunosorbent assays, high-performance liquid chromatography (HPLC) with electrochemical detection (ECD) or liquid chromatography tandem mass spectrometry (LCMS/MS). Since the first two methods are either extremely laborious, necessitate low sample run numbers, result in slow turnaround times or are subject to analytical interference, a robust, routine clinical method is not achievable. We established a novel sample preparation method to measure plasma free metadrenalines using LCMS/MS.
Methods: Three different solid-phase extraction (SPE) methods were compared: hydrophilic-lipophilic balance sorbent (HLB), weak cation exchange (WCX) and mixed mode cation exchange (MCX) and their ability to remove interfering compounds prior to LCMS/MS analysis. Maximum recovery of plasma free metadrenaline and plasma free normetadrenaline were achieved by positively charging compounds prior to SPE application.
Results: Compared with HLB and WCX cartridges, MCX extraction resulted in chromatography without co-eluting interference with superior assay precision and accuracy. Additionally, samples that could not be quantified because of interference using HPLC/ECD could be readily assayed using this new method.
Conclusions: The use of the MCX SPE method with LCMS/MS detection provides an improved assay to measure plasma free metadrenalines in comparison to many available alternative methods.