The granulocyte-macrophage colony stimulating factor (GM-CSF) is a cytokine that regulates the proliferation and differentiation of hematopoietic cells and activates granulocytes and macrophages. A reversed-phase liquid chromatography (RP-LC) method was validated for the assessing of the stability of non-glycosylated recombinant rhGM-CSF (Molgramostim) in biopharmaceutical formulations. The RP-LC method was carried out on a Jupiter C(4) column (250 mm × 4.6 mm i.d.), maintained at 45 °C. The mobile phase A consisted of 0.1% TFA and the mobile phase B was acetonitrile with 0.1% TFA in acetonitrile, run at a flow rate of 1 mL/min, and using photodiode array (PDA) detection at 214 nm. Chromatographic separation was obtained with a retention time of 29.2 min, and was linear over the concentration range of 2-300 μg/mL (r(2) = 0.9992). Specificity was established in degradation studies. Moreover, the in vitro cytotoxicity test of the degraded products showed significant differences (p < 0.05). The method was applied to the assessment of rhGM-CSF and related proteins in biopharmaceutical dosage forms, and the results were correlated to those of a bioassay. It is concluded that the employment of RP-LC in conjunction with current methods allows a great improvement in monitoring stability, quality control and thereby assures the therapeutic efficacy.
Copyright © 2011 The International Alliance for Biologicals. Published by Elsevier Ltd. All rights reserved.