Erythrocyte sedimentation rate measurement by VES Matic Cube 80 in relation to inflammation plasma proteins

J Clin Lab Anal. 2011;25(3):198-202. doi: 10.1002/jcla.20457.

Abstract

Westergren method is considered as the reference procedure to measure Erythrocyte Sedimentation Rate (ESR) by the International Council for Standardization in Haematology. However, a closed automated method, VES Matic Cube 80 (DIESSE S.p.A., Siena, Italy), has been introduced as a new ESR measurement instrument. In this article, we report two different studies: first, we compared the two methods (Westergren and VES Matic Cube 80) and second, we correlated the inflammatory state of 248 patients with their ESR values. Total protein, albumin, C-reactive protein, and other inflammatory proteins were detected in each sample. The results obtained using VES Matic Cube 80 demonstrated a good correlation with those obtained using the Westergren method (Ordinary linear regression: y=0.955x-0.205, r(2) =0.816, P<0.05; Passing-Bablock regression equation: y=0.9153x-0.5763; Bland-Altman analysis: bias 1.2; limits of agreement -17.4-19.9) and with the inflammatory protein levels (CRP: r=0.554 and r=0.498 and Fibrinogen: r=0.699 and r=0.663 for Ves Matic Cube 80 and Westergren, respectively), supporting the hypothesis that VES Matic Cube 80 offers a fast and safe ESR determination, ensuring precision and a very good correlation with the reference method.

MeSH terms

  • Blood Proteins / metabolism*
  • Blood Sedimentation
  • Female
  • Hematologic Tests / instrumentation*
  • Hematologic Tests / methods*
  • Humans
  • Inflammation / blood*
  • Linear Models
  • Male
  • Middle Aged

Substances

  • Blood Proteins