Dynamic interaction between melamine and cyanuric acid in artificial urine investigated by quartz crystal microbalance

Analyst. 2011 Jun 21;136(12):2482-8. doi: 10.1039/c1an15119c. Epub 2011 May 11.

Abstract

The dynamic interaction between melamine (M) and cyanuric acid (CA) in artificial urine was investigated by a flow injection analysis-quartz crystal microbalance (FIA-QCM) system. Melamine was used as the recognition element and immobilized onto the QCM gold surface. The process of M and CA interaction was recorded by FIA-QCM in real-time. The multilayer complex of M and CA was successfully formed on the crystal surface of the QCM when CA and M were introduced into the FIA-QCM system alternately. The influence of pH on the M and CA interaction indicated that the M-CA multilayer complex possesses high stability over a wide range of pH values, especially in physiological urine condition. The association and dissociation rate constants were determined under artificial urine condition and the association constant for M and CA was calculated as 10(2) L mol(-1). The M-CA complex was further characterized by Fourier transform infrared spectroscopy (FT-IR) and atomic force microscopy (AFM). Mechanism study suggested that the formation of the multilayer complex in artificial urine was mainly attributed to the intermolecular hydrogen bonding between M and CA. The work provides direct insight into the dynamic interaction between M and CA under physiological conditions and would be helpful for the research of kidney stones induced by melamine contamination.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Flow Injection Analysis
  • Gold / chemistry
  • Hydrogen Bonding
  • Hydrogen-Ion Concentration
  • Microscopy, Atomic Force / methods
  • Quartz Crystal Microbalance Techniques / methods*
  • Spectroscopy, Fourier Transform Infrared / methods
  • Triazines / chemistry*
  • Triazines / urine

Substances

  • Triazines
  • Gold
  • cyanuric acid
  • melamine