Registration accuracy of three-dimensional surface and cone beam computed tomography data for virtual implant planning

Clin Oral Implants Res. 2012 Apr;23(4):447-52. doi: 10.1111/j.1600-0501.2011.02159.x. Epub 2011 Apr 13.

Abstract

Objective: Virtual wax-ups based on three-dimensional (3D) surface models can be matched (i.e. registered) to cone beam computed tomography (CBCT) data of the same patient for dental implant planning. Thereby, implant planning software can visualize anatomical and prosthetic information simultaneously. The aim of this study is to assess the accuracy of a newly developed registration process.

Material and methods: Data pairs of CBCT and 3D surface data of 16 patients for dental implant planning were registered and the discrepancy between the visualized 3D surface data and the corresponding CBCT data were measured on 64 teeth at seven points by two investigators in two iterations with a total of 1792 measurements.

Results: All data pairs were matched successfully and mean distances between CBCT and 3D surface data were between 0.03(±0.33) and 0.14(±0.18) mm. At two of seven measuring points, statistically significant correlations were determined between the measured error and the presence and type of restorations. Registration errors in maxilla and mandible were not statistically significantly different.

Conclusion: According to the results of this study, registration of 3D surface data and CBCT data works reliably and is sufficiently accurate for dental implant planning. Thereby, barium-sulfate scanning templates can be avoided and dental implant planning can be accomplished fully virtual.

MeSH terms

  • Computer-Aided Design*
  • Cone-Beam Computed Tomography*
  • Dental Implantation, Endosseous*
  • Humans
  • Imaging, Three-Dimensional*
  • Models, Dental
  • Patient Care Planning
  • Reproducibility of Results
  • Retrospective Studies
  • Software
  • Statistics, Nonparametric
  • Surgery, Computer-Assisted*
  • User-Computer Interface*