High resolution spectroscopic imaging of GABA at 3 Tesla

Magn Reson Med. 2011 Mar;65(3):603-9. doi: 10.1002/mrm.22671. Epub 2010 Nov 4.

Abstract

A spin echo-based MRSI sequence was developed to acquire edited spectra of γ-aminobutyric acid in an entire slice. Water and lipid signals were suppressed by a dual-band presaturation sequence, which included integrated outer volume suppression pulses for additional lipid suppression. Experiments in three normal volunteers were performed at 3 T using a 32-channel head coil. High signal-to-noise ratio spectra and metabolic images of γ-aminobutyric acid were acquired from nominal 4.5 cm3 voxels (estimated actual voxel size 7.0 cm3) in a scan time of 17 min. The sequence is also expected to co-edit homocarnosine and macromolecules, giving a composite γ-aminobutyric acid+ resonance. The γ-aminobutyric acid+ to water ratio was measured using a companion water MRSI scan and was found to correlate linearly with the % gray matter (GM) of each voxel (γ-aminobutyric acid+/water=(1.5×GM+3.2)×10(-5), R=0.27), with higher γ-aminobutyric acid+ levels in gray matter compared with white. In conclusion, high signal-to-noise ratio γ-aminobutyric acid-MRSI is possible at 3 T within clinically feasible scan times.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Algorithms*
  • Brain Chemistry*
  • Echo-Planar Imaging / methods*
  • Female
  • Humans
  • Magnetic Resonance Spectroscopy / methods*
  • Male
  • Tissue Distribution
  • gamma-Aminobutyric Acid / analysis*

Substances

  • gamma-Aminobutyric Acid