Closed-loop insulin delivery during pregnancy complicated by type 1 diabetes

Diabetes Care. 2011 Feb;34(2):406-11. doi: 10.2337/dc10-1796. Epub 2011 Jan 7.

Abstract

Objective: This study evaluated closed-loop insulin delivery with a model predictive control (MPC) algorithm during early (12-16 weeks) and late gestation (28-32 weeks) in pregnant women with type 1 diabetes.

Research design and methods: Ten women with type 1 diabetes (age 31 years, diabetes duration 19 years, BMI 24.1 kg/m(2), booking A1C 6.9%) were studied over 24 h during early (14.8 weeks) and late pregnancy (28.0 weeks). A nurse adjusted the basal insulin infusion rate from continuous glucose measurements (CGM), fed into the MPC algorithm every 15 min. Mean glucose and time spent in target (63-140 mg/dL), hyperglycemic (>140 to ≥ 180 mg/dL), and hypoglycemic (<63 to ≤ 50 mg/dL) were calculated using plasma and sensor glucose measurements. Linear mixed-effects models were used to compare glucose control during early and late gestation.

Results: During closed-loop insulin delivery, median (interquartile range) plasma glucose levels were 117 (100.8-154.8) mg/dL in early and 126 (109.8-140.4) mg/dL in late gestation (P = 0.72). The overnight mean (interquartile range) plasma glucose time in target was 84% (50-100%) in early and 100% (94-100%) in late pregnancy (P = 0.09). Overnight mean (interquartile range) time spent hyperglycemic (>140 mg/dL) was 7% (0-40%) in early and 0% (0-6%) in late pregnancy (P = 0.25) and hypoglycemic (<63 mg/dL) was 0% (0-3%) and 0% (0-0%), respectively (P = 0.18). Postprandial glucose control, glucose variability, insulin infusion rates, and CGM sensor accuracy were no different in early or late pregnancy.

Conclusions: MPC algorithm performance was maintained throughout pregnancy, suggesting that overnight closed-loop insulin delivery could be used safely during pregnancy. More work is needed to achieve optimal postprandial glucose control.

Publication types

  • Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Algorithms
  • Biosensing Techniques / methods
  • Biosensing Techniques / statistics & numerical data
  • Blood Glucose / drug effects
  • Blood Glucose / metabolism
  • Circadian Rhythm
  • Diabetes Mellitus, Type 1 / diagnosis
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Female
  • Humans
  • Hyperglycemia / drug therapy
  • Hypoglycemic Agents / administration & dosage*
  • Insulin / administration & dosage*
  • Insulin Infusion Systems* / standards
  • Pregnancy
  • Pregnancy in Diabetics / diagnosis
  • Pregnancy in Diabetics / drug therapy*
  • Reproducibility of Results

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin

Associated data

  • ISRCTN/ISRCTN62568875