Comparison of the RealTime HIV-1, COBAS TaqMan 48 v1.0, Easy Q v1.2, and Versant v3.0 assays for determination of HIV-1 viral loads in a cohort of Canadian patients with diverse HIV subtype infections

J Clin Microbiol. 2011 Jan;49(1):118-24. doi: 10.1128/JCM.00685-10. Epub 2010 Nov 17.

Abstract

HIV clinics in Canada provide care to an increasing number of patients born outside of Canada with HIV-1 non-B subtype infections. Because the Easy Q HIV-1 v1.2 assay (EQ; bioMérieux) failed to detect some non-B subtype infections, a multiassay HIV-1 viral load (VL) study was conducted with patients with diverse HIV subtype infections. Patients were enrolled from the Southern Alberta HIV Clinic (SAC), Calgary, Alberta, Canada (n = 349) and the McGill HIV Clinic (MHC), Montreal, Quebec, Canada (n = 20) and had four or five tubes of blood drawn for testing by EQ and three other commercial HIV VL assays: (i) the Versant 3.0 HIV-1 test, with the Versant 440 instrument (branched DNA [bDNA]; Siemens), (ii) the RealTime HIV-1 test, with the m2000rt instrument (m2000rt; Abbott Molecular Diagnostics), and (iii) the COBAS AmpliPrep TaqMan HIV-1 48 test (CAP-CTM; Roche Molecular Diagnostics). Blood was processed according to the individual manufacturer's requirements and stored frozen at -86°C. The HIV subtype was known for patients who had undergone HIV genotypic resistance testing (Virco, Belgium). Data analyses were done using standard statistical methods within Stata 9.0 (StataCorp, College Station, TX). A total of 371 samples were tested on 369 patients, of whom 291 (81%) had a Virco genotype result of B (195; 53%) or non-B (96; 26%) subtypes A to D and F to K, as well as circulating recombinant forms (CRFs) (i.e., CRF01_AE and CRF02_AG). Most (58/78; 74%) patients of unknown subtype were recent African emigrants who likely have non-subtype B infection. Overall bias was small in pairwise Bland-Altman plots, but the limits of agreement between assays were wide. Discordant viral load results occurred for 98 samples and were due to missing values, false negatives, and significant underquantification that varied by HIV subtype. Results were obtained for all 371 samples with m2000rt, but for only 357 (97%) with CAP-CTM, 338 (92%) with EQ, and 276 (75%) with bDNA due to errors/equipment failures. False-negative results (nondetection of viral RNA versus other assay results) occurred for all platforms, as follows: for m2000rt, 8 (2%) [B(4) and non-B(4) subtypes], CAP-CTM, 9 (2.5%) [B(6) and non-B(3) subtypes]; EQ, 20 (6%) [B(7) and non-B(13) subtypes]; bDNA, 5 (2%) [B(1) and C(4)]. EQ and bDNA had the highest rates of underquantification by ≥ 1.0 log(10) copies/ml, mainly for HIV non-B subtypes. Performance significantly varied between HIV VL platforms according to subtype. HIV viral diversity in the population being tested must be considered in selection of the viral load platform.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Alberta
  • Blood / virology
  • Female
  • Genetic Variation*
  • HIV Infections / virology*
  • HIV-1 / classification
  • HIV-1 / genetics
  • HIV-1 / isolation & purification*
  • Humans
  • Male
  • Middle Aged
  • Molecular Diagnostic Techniques / methods*
  • Quebec
  • Reagent Kits, Diagnostic
  • Viral Load / methods*

Substances

  • Reagent Kits, Diagnostic