Solid core column technology applied to HPLC-FD of paralytic shellfish toxins

Toxicon. 2011 Jan;57(1):179-82. doi: 10.1016/j.toxicon.2010.11.004. Epub 2010 Nov 13.

Abstract

Pre-column oxidation liquid chromatography with fluorescence detection is a chemical method for analyzing paralytic shellfish toxins. In order to improve the sample throughput and efficiency of AOAC Method 2005.06, solid core particle column technology was evaluated. We demonstrate that supplanting the original fully porous particle column with a solid core particle column reduces sample analysis time from 15 to 5 min per sample and improves resolution.

MeSH terms

  • Animals
  • Chromatography, High Pressure Liquid
  • Food Contamination / analysis*
  • Harmful Algal Bloom
  • High-Throughput Screening Assays
  • Marine Toxins / analysis*
  • Phytoplankton
  • Shellfish / analysis*
  • Shellfish Poisoning
  • Solid Phase Extraction / methods

Substances

  • Marine Toxins