A coaxial thermocouple for shock tunnel applications

Rev Sci Instrum. 2010 Oct;81(10):104905. doi: 10.1063/1.3494605.

Abstract

A chromel-constantan coaxial surface junction thermocouple has been designed, fabricated, calibrated, and tested to measure the temperature-time history on the surface of a body in a hypersonic freestream of Mach 8 in a shock tunnel. The coaxial thermocouple with a diameter of 3.25 mm was flush mounted in the surface of a hemisphere of 25 mm diameter. The hypersonic freestream was of a very low temperature and density, and had a flow time of about a millisecond. Preliminary test results indicate that the thermocouple is quite sensitive to low temperature-rarefied freestreams, and also has a response time of a few microseconds (≈5 μs) to meet the requirements of short duration transient measurements. The sensor developed is accurate, robust, reproducible, and is highly inexpensive.