Generalized MR interferography

Magn Reson Med. 1990 Dec;16(3):390-402. doi: 10.1002/mrm.1910160305.

Abstract

Generalized MR interferography allows the measurement of parameters which affect the phase of the MR signal. Interference is generated by the simultaneous generation of two echoes in one excitation cycle. Local variations of the signal phase are transformed into displacement of the stripe pattern generated by interference. A number of possible experiments based on this principle are presented. The resulting interferographic images allow the fast determination of parameters such as field inhomogeneity, susceptibility, flow, and motion.

MeSH terms

  • Fourier Analysis
  • Humans
  • Interferometry / methods
  • Magnetic Resonance Imaging / methods*
  • Models, Structural