Performance of external and internal coil configurations for prostate investigations at 7 T

Magn Reson Med. 2010 Dec;64(6):1625-39. doi: 10.1002/mrm.22552. Epub 2010 Aug 25.

Abstract

Three different coil configurations were evaluated through simulation and experimentally to determine safe operating limits and evaluate subject size-dependent performance for prostate imaging at 7 T. The coils included a transceiver endorectal coil (trERC), a 16-channel transceiver external surface array (trESA) and a trESA combined with a receive-only ERC (trESA+roERC). Although the transmit B(1) (B 1+) homogeneity was far superior for the trESA, the maximum achievable B 1+ is subject size dependent and limited by transmit chain losses and amplifier performance. For the trERC, limitations in transmit homogeneity greatly compromised image quality and limited coverage of the prostate. Despite these challenges, the high peak B 1+ close to the trERC and subject size-independent performance provides potential advantages especially for spectroscopic localization where high-bandwidth radiofrequency pulses are required. On the receive side, the combined trESA+roERC provided the highest signal-to-noise ratio and improved homogeneity over the trERC resulting in better visualization of the prostate and surrounding anatomy. In addition, the parallel imaging performance of the trESA+roERC holds strong promise for diffusion-weighted imaging and dynamic contrast-enhanced MRI.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Equipment Design
  • Humans
  • Image Enhancement / instrumentation
  • Image Enhancement / methods
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Phantoms, Imaging
  • Prostate / anatomy & histology*
  • Rectum