Reproducibility of peripapillary retinal nerve fiber layer thickness and optic nerve head parameters measured with cirrus HD-OCT in glaucomatous eyes

Invest Ophthalmol Vis Sci. 2010 Nov;51(11):5724-30. doi: 10.1167/iovs.10-5222. Epub 2010 Jun 23.

Abstract

Purpose: To assess the reproducibility of peripapillary retinal nerve fiber layer (RNFL) thickness and optic nerve head (ONH) parameters measured with Cirrus HD-OCT in glaucomatous eyes.

Methods: Fifty-five glaucomatous eyes were included in the study. The optic disc cube 200 × 200 protocol was used to obtain three scans during the same visit to evaluate the intravisit reproducibility. One scan on 4 additional days within a 2-month period of the first session was obtained to assess intervisit reproducibility. Intraclass correlation coefficient (ICC), coefficient of variation (CV), and test-retest SD (TRT SD) were calculated for each RNFL and ONH parameter. The formula 1.645 × √2 × intervisit TRT SD provides an upper tolerance limit to variability beyond which nonphysiologic change should be considered.

Results: All ICCs were excellent, ranging from 83.9% to 99.2% for intravisit measurements and from 80.8% to 99.1% for intervisit measurements. Cup/disc area ratio had the lowest CV (1.1%) in either type of measurement, followed by average RNFL thickness (1.9% and 2.7%). Nasal clock hours and quadrants showed the poorest reproducibility as did the clock hour directly temporally. The intervisit tolerance limit for average RNFL thickness was 3.89 μm.

Conclusions: Intravisit and intervisit measurements of peripapillary RNFL thickness and ONH parameters with Cirrus HD-OCT showed excellent reproducibility, indicating that this instrument may be useful in monitoring glaucoma progression. When comparing two measurements from the same eye on two different visits, a reproducible decrease in average RNFL thickness of approximately 4 μm or more may be considered a statistically significant change from baseline.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Disease Progression
  • Female
  • Glaucoma / diagnosis*
  • Glaucoma / physiopathology
  • Humans
  • Intraocular Pressure
  • Male
  • Middle Aged
  • Nerve Fibers / pathology*
  • Optic Disk / pathology*
  • Optic Nerve Diseases / diagnosis*
  • Optic Nerve Diseases / physiopathology
  • Reproducibility of Results
  • Retinal Ganglion Cells / pathology*
  • Tomography, Optical Coherence*