Far-infrared high-resolution Fourier transform spectrometer

Appl Opt. 1987 Sep 15;26(18):3818-22. doi: 10.1364/AO.26.003818.

Abstract

Design and performance of a high-resolution Fourier transform spectrometer for laboratory molecular spectroscopy in the 8-200-cm(-1) region are discussed. A folding of the beam path is used to obtain the maximum path difference of 4 m with a mirror stroke of 1 m. The measured linewidth of 0.0019 cm(-1) is in agreement with the expected theoretical resolution.