Monte carlo study of MOSFET packaging, optimised for improved energy response: single MOSFET filtration

Radiat Prot Dosimetry. 2010 Sep;141(1):10-7. doi: 10.1093/rpd/ncq144. Epub 2010 May 11.

Abstract

Monte Carlo simulations of the energy response of a conventionally packaged single metal-oxide field effect transistors (MOSFET) detector were performed with the goal of improving MOSFET energy dependence for personal accident or military dosimetry. The MOSFET detector packaging was optimised. Two different 'drop-in' design packages for a single MOSFET detector were modelled and optimised using the GEANT4 Monte Carlo toolkit. Absorbed photon dose simulations of the MOSFET dosemeter placed in free-air response, corresponding to the absorbed doses at depths of 0.07 mm (D(w)(0.07)) and 10 mm (D(w)(10)) in a water equivalent phantom of size 30 x 30 x 30 cm(3) for photon energies of 0.015-2 MeV were performed. Energy dependence was reduced to within + or - 60 % for photon energies 0.06-2 MeV for both D(w)(0.07) and D(w)(10). Variations in the response for photon energies of 15-60 keV were 200 and 330 % for D(w)(0.07) and D(w)(10), respectively. The obtained energy dependence was reduced compared with that for conventionally packaged MOSFET detectors, which usually exhibit a 500-700 % over-response when used in free-air geometry.