High-resolution photoacoustic imaging of ocular tissues

Ultrasound Med Biol. 2010 May;36(5):733-42. doi: 10.1016/j.ultrasmedbio.2010.02.006.

Abstract

Optical coherence tomography (OCT) and ultrasound (US) are methods widely used for diagnostic imaging of the eye. These techniques detect discontinuities in optical refractive index and acoustic impedance, respectively. Because these both relate to variations in tissue density or composition, OCT and US images share a qualitatively similar appearance. In photoacoustic imaging (PAI), short light pulses are directed at tissues, pressure is generated due to a rapid energy deposition in the tissue volume and thermoelastic expansion results in generation of broadband US. PAI thus depicts optical absorption, which is independent of the tissue characteristics imaged by OCT or US. Our aim was to demonstrate the application of PAI in ocular tissues and to do so with lateral resolution comparable to OCT. We developed two PAI assemblies, both of which used single-element US transducers and lasers sharing a common focus. The first assembly had optical and 35-MHz US axes offset by a 30 degrees angle. The second assembly consisted of a 20-MHz ring transducer with a coaxial optics. The laser emitted 5-ns pulses at either 532 nm or 1064 nm, with spot sizes at the focus of 35 microm for the angled probe and 20 microm for the coaxial probe. We compared lateral resolution by scanning 12.5 microm diameter wire targets with pulse/echo US and PAI at each wavelength. We then imaged the anterior segment in whole ex vivo pig eyes and the choroid and ciliary body region in sectioned eyes. PAI data obtained at 1064 nm in the near infrared had higher penetration but reduced signal amplitude compared to that obtained using the 532 nm green wavelength. Images were obtained of the iris, choroid and ciliary processes. The zonules and anterior cornea and lens surfaces were seen at 532 nm. Because the laser spot size was significantly smaller than the US beamwidth at the focus, PAI images had superior resolution than those obtained using conventional US.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Elasticity Imaging Techniques / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Eye / diagnostic imaging*
  • In Vitro Techniques
  • Ophthalmoscopes*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Swine
  • Transducers*