Microwave ablation: results with double 915 MHz antennae in ex vivo bovine livers

Eur J Radiol. 2011 Aug;79(2):214-7. doi: 10.1016/j.ejrad.2010.03.015. Epub 2010 Apr 14.

Abstract

Purpose: To investigate the settings for the optimal microwave ablation geometry with the simultaneous application of double 915 MHz antennae in ex vivo bovine livers, so as to provide the technical basis for treating large liver tumor in one ablation session.

Materials and methods: MWAs were performed on ex vivo bovine livers by simultaneously application of double 915 MHz internally cooled-shaft antennae. Four power settings (50, 60, 70 and 80 W) were used during MWAs, while application time was fixed at 10 min. Three inter-antenna distances (2.0, 2.5 and 3.0 cm) were used. Diameters and shapes of the coagulation zones were observed on gross specimens.

Results: (1) The coagulation shape was related to the inter-antenna distance, which was most spherical at an inter-antenna distance of 2.0 cm. A recess of the coagulation zone was observed at an inter-antenna distances of 2.5 and 3.0 cm. (2) The long-axis and short-axis coagulation diameter enlarged with increasing power output. However, there were no significant differences in the coagulation diameters between 70 and 80 W (P>0.05). More desirable coagulation geometry could be obtained by simultaneous application of double antennae at 70 W for 10 min with an inter-antenna distance of 2.0 cm, the long-axis and short-axis coagulation diameter were 6.95 ± 0.32 cm and 5.30 ± 0.22 cm, respectively.

Conclusion: Simultaneous application of double 915 MHz antennae can generate large coagulation zones with desirable shape which may be advantageous for treating large liver tumor in one ablation session.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Catheter Ablation / instrumentation*
  • Cattle
  • Equipment Design
  • In Vitro Techniques
  • Liver / surgery*
  • Microwaves / therapeutic use*