Stability of medial locking plate and compression screw versus two crossed screws for lapidus arthrodesis

Foot Ankle Int. 2010 Feb;31(2):158-63. doi: 10.3113/FAI.2010.0158.

Abstract

Background: Lapidus (first metatarsocuneiform joint) arthrodesis is an established procedure for the management of hallux valgus. This study investigated the utility of fixation with a medial locking plate with adjunct compression screw versus fixation with two crossed screws.

Materials and methods: Eight pairs of fresh-frozen human specimens were used in a matched pair test. Bone mineral density (BMD) was measured with peripheral quantitative computed tomography (pQCT). Fixation with two 4-mm-diameter crossed screws was compared versus a medial locking plate (X-Locking Plate 2.4/2.7; Synthes, Solothurn, Switzerland) with adjunct 4-mm-diameter compression screw. The specimens were tested in a four-point bending test. Parameters obtained were initial stiffness; plantar joint-line gapping after one cycle, 100 and 1000 cycles; and number of cycles to failure. Failure was defined as more than or equal to 3 mm plantar gapping.

Results: The groups did not differ significantly with regard to BMD (p = 0.866) and initial stiffness (p = 0.889). The plate-and-screw construct showed significantly less movement during testing, and significantly (p = 0.001) more cycles to failure than did the crossed-screw construct. There was a significant correlation (crossed-screw construct: p = 0.014; plate-and-screw construct: p = 0.010) between BMD and the number of cycles to failure.

Conclusion: Under cyclic loading conditions, the construct using a medial locking plate with adjunct compression screw was superior to the construct using two crossed screws.

Clinical relevance: The medial locking-plate technique described could help shorten the period of nonweightbearing and reduce the risk of non-union.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Arthrodesis / instrumentation*
  • Biomechanical Phenomena
  • Bone Density
  • Bone Plates*
  • Bone Screws*
  • Cadaver
  • Female
  • Hallux Valgus / diagnostic imaging
  • Hallux Valgus / surgery*
  • Humans
  • Male
  • Middle Aged
  • Statistics, Nonparametric
  • Tomography, X-Ray Computed