Preoperative planning of total hip replacement using the TraumaCad™ system

Arch Orthop Trauma Surg. 2010 Dec;130(12):1429-32. doi: 10.1007/s00402-010-1046-y. Epub 2010 Jan 13.

Abstract

Introduction: Templating is now the standard approach for preoperative planning of total joint replacement and fracture fixation. The aim of this study was to assess the accuracy of new software, TraumaCad™, for preoperative planning for total hip replacement.

Methods: This software enables the import and export of all picture archiving communication system (PACS) files from local working stations. It can integrate with all PACS systems. We retrospectively reviewed TraumaCad™-generated templates of digitalized preoperative radiographs of 73 consecutive patients.

Results: The acetabular component, measured within ± 1 size, was accurate in 65 patients (89%), and the femoral stem design component was accurate in 70 patients (97%). The accuracy of the femoral size was predicted in 57 patients within 1 size (87%) and in 65 patients within 2 sizes (96%) out of 68 patients in the Corail femoral component group. TraumaCad™ successfully predicts the sizes of femoral and acetabular component and easily integrates with all PACS files.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / methods*
  • Decision Making, Computer-Assisted*
  • Female
  • Femur / diagnostic imaging
  • Hip Joint / diagnostic imaging
  • Humans
  • Male
  • Middle Aged
  • Patient Care Planning*
  • Radiography
  • Retrospective Studies
  • Software*