Opening of the blood-brain barrier by low-frequency (28-kHz) ultrasound: a novel pinhole-assisted mechanical scanning device

Ultrasound Med Biol. 2010 Feb;36(2):325-35. doi: 10.1016/j.ultrasmedbio.2009.10.004. Epub 2009 Dec 16.

Abstract

Disruption of the blood-brain barrier (BBB) may be transiently achieved via high-frequency focused spherical ultrasound in the presence of microbubbles. In this experimental animal study, we sought to determine whether focal reversible opening of the BBB may be achieved using low-frequency (i.e., 20-30 kHz) planar ultrasonic waves. In the presence of microbubbles, we were able to obtain BBB opening using non-focused ultrasound irradiation with a frequency as low as 28 kHz. We also achieved a tight regulation of the ultrasound patterns by using a mechanical scanning device equipped with a pinhole. Histologic examination of the brains supported the feasibility of our system. The areas of BBB disruption obtained with this method were large enough to cover a typical circumscribed cerebral tumor mass. The inherent advantages of our BBB opening method include an improved portability, the possibility to obtain fairly wide areas of BBB opening and a low incidence of hemorrhagic complications. In addition, our system has the potential to reduce the need for image guidance for treating superficial brain lesions.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Blood-Brain Barrier / diagnostic imaging*
  • Brain / pathology*
  • Brain Neoplasms / pathology
  • Contrast Media
  • Disease Models, Animal
  • Microbubbles*
  • Rats
  • Reference Standards
  • Ultrasonic Therapy / methods*
  • Ultrasonic Therapy / standards
  • Ultrasonics*
  • Ultrasonography

Substances

  • Contrast Media