Daily image guidance with cone-beam computed tomography for head-and-neck cancer intensity-modulated radiotherapy: a prospective study

Int J Radiat Oncol Biol Phys. 2010 Apr;76(5):1353-9. doi: 10.1016/j.ijrobp.2009.03.059. Epub 2009 Jun 18.

Abstract

Purpose: To report on a prospective clinical trial of the use of daily kilovoltage cone-beam computed tomography (CBCT) to evaluate the interfraction and residual error motion of patients undergoing intensity-modulated radiotherapy for head-and-neck cancer.

Methods and materials: Patients were treated with intensity-modulated radiotherapy with an Elekta linear accelerator using a mounted CBCT scanner. CBCT was performed before every treatment, and translational (but not rotational) corrections were performed. At least once per week, a CBCT scan was obtained after intensity-modulated radiotherapy. Variations were measured in the medial-lateral, superoinferior, and anteroposterior dimensions, as well as in the rotation around these axes.

Results: A total of 28 consecutive patients (1,013 CBCT scans) were studied. The average interfraction shift was 1.4 +/- 1.4, 1.7 +/- 1.9, and 1.8 +/- 2.1 mm in the medial-lateral, superoinferior, and anteroposterior dimensions, respectively. The corresponding average residual error shifts were 0.7 +/- 0.8, 0.9 +/- 0.9, and 0.9 +/- 0.9 mm. These data indicate that in the absence of daily CBCT image-guided radiotherapy, a clinical target volume to planning target volume margin of 3.9, 4.1, and 4.9 mm is needed in the medial-lateral, superoinferior, and anteroposterior dimensions, respectively. With daily CBCT, corresponding margins of 1.6, 2.5, and 1.9 mm should be acceptable. Subgroup analyses showed that larynx cancers and/or intratreatment weight loss indicate a need for slightly larger clinical target volume to planning target volume margins.

Conclusion: The results of our study have shown that image-guided radiotherapy using CBCT for head-and-neck cancer is effective. These data suggest it allows a reduction in the clinical target volume to planning target volume margins by about 50%, which could facilitate future studies of dose escalation and/or improved toxicity reduction. Caution is particularly warranted for cases in which the targets are mobile (e.g., the tongue).

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Calibration
  • Cone-Beam Computed Tomography*
  • Dose Fractionation, Radiation
  • Feasibility Studies
  • Female
  • Head and Neck Neoplasms / diagnostic imaging*
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Male
  • Middle Aged
  • Movement*
  • Prospective Studies
  • Radiotherapy, Intensity-Modulated / methods*