A randomized, controlled, multicenter trial to evaluate the safety and efficacy of zotarolimus- versus paclitaxel-eluting stents in de novo occlusive lesions in coronary arteries The ZoMaxx I trial

JACC Cardiovasc Interv. 2008 Oct;1(5):524-32. doi: 10.1016/j.jcin.2008.08.010.

Abstract

Objectives: A novel zotarolimus-eluting coronary stent system (ZoMaxx, Abbott Laboratories, Abbott Park, Illinois) was compared with a paclitaxel-eluting coronary stent (Taxus Express2) in a randomized trial of percutaneous intervention for de novo coronary artery stenosis. The primary end point was defined as noninferiority of in-segment late lumen loss after 9 months.

Background: The ZoMaxx stent system elutes 10 microg/mm zotarolimus using a phosphorylcholine polymer loaded onto a novel stainless steel stent platform containing a 0.0007-inch inner layer of tantalum.

Methods: Twenty-nine investigative sites in Europe, Australia, and New Zealand enrolled 401 patients, 396 of whom received a study stent.

Results: After 9 months, late lumen loss was significantly greater in the ZoMaxx group (in-stent 0.67 +/- 0.57 mm vs. 0.45 +/- 0.48 mm; p < 0.001; in-segment 0.43 +/- 0.60 mm vs. 0.25 +/- 0. 45 mm; p = 0.003), resulting in significantly higher rates of >50% angiographic restenosis (in-stent 12.9% vs. 5.7%; p = 0.03; in-segment 16.5% vs. 6.9%; p = 0.007). The upper bound of the 95% confidence interval on the difference in in-segment late lumen loss between the 2 treatment groups (0.27 mm) exceeded the 0.25 mm value pre-specified for noninferiority. There were no significant differences between ZoMaxx and Taxus-treated groups with respect to target lesion revascularization (8.0% vs. 4.1%; p = 0.14), major adverse cardiac events (12.6% vs. 9.6%; p = 0.43), or stent thrombosis (0.5% in both groups).

Conclusions: After 9 months, the ZoMaxx stent showed less neointimal inhibition than the Taxus stent, as shown by higher in-stent late loss and restenosis by qualitative coronary angiography.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Australia
  • Cardiovascular Agents / administration & dosage*
  • Coronary Angiography
  • Coronary Restenosis / etiology
  • Coronary Restenosis / pathology
  • Coronary Restenosis / prevention & control*
  • Coronary Stenosis / pathology
  • Coronary Stenosis / therapy*
  • Drug-Eluting Stents*
  • Europe
  • Female
  • Humans
  • Logistic Models
  • Male
  • Middle Aged
  • New Zealand
  • Paclitaxel / administration & dosage*
  • Prospective Studies
  • Prosthesis Design
  • Risk Assessment
  • Severity of Illness Index
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Thrombosis / etiology
  • Thrombosis / pathology
  • Thrombosis / prevention & control*
  • Time Factors
  • Treatment Outcome
  • Ultrasonography, Interventional

Substances

  • Cardiovascular Agents
  • zotarolimus
  • Paclitaxel
  • Sirolimus