CT fluoroscopy-guided biopsy of 1,000 pulmonary lesions performed with 20-gauge coaxial cutting needles: diagnostic yield and risk factors for diagnostic failure

Chest. 2009 Dec;136(6):1612-1617. doi: 10.1378/chest.09-0370. Epub 2009 May 8.

Abstract

Background: Although conventional CT scan-guided needle biopsy is an established diagnostic method for pulmonary lesions, few large studies have been conducted on the diagnostic outcomes of CT fluoroscopy-guided lung biopsy. We have conducted a retrospective analysis to evaluate the diagnostic outcomes of 1,000 CT fluoroscopy-guided lung biopsies performed with 20-gauge coaxial cutting needles.

Methods: We determined the diagnostic yield of CT fluoroscopy-guided lung biopsies performed with 20-gauge coaxial cutting needles for 1,000 lesions in 901 patients. Independent risk factors for diagnostic failure (ie, nondiagnostic, false-positive, and false-negative results) were determined with multivariate logistic regression analysis.

Results: The biopsy results were nondiagnostic in 0.6% of the lesions (6 of 1,000 lesions). The sensitivity and specificity for the diagnosis of malignancy was 94.2% (741 of 787 lesions) and 99.1% (211 of 213 lesions), respectively; diagnostic accuracy was 95.2% (952 of 1,000 lesions). For lesions measuring <or= 1.0 cm, the diagnostic accuracy was 92.7% (140 of 151 lesions). The significant independent risk factors for diagnostic failure were as follows: the acquisition of two or fewer specimens (odds ratio [OR], 2.43; p = 0.007), lesions in the lower lobe (OR, 2.50; p = 0.003), malignant lesions (OR, 7.16; p = 0.007), and lesions measuring <or= 1.0 cm (OR, 3.85; p = 0.016) and >or= 3.1 cm (OR, 4.32; p = 0.007).

Conclusions: CT fluoroscopy-guided lung biopsy performed with 20-gauge coaxial cutting needles resulted in a high diagnostic yield, even in the case of small lesions. Factors such as the acquisition of two or fewer specimens, lesions in the lower lobe, malignant lesions, and lesions measuring <or= 1.0 cm or >or= 3.1 cm significantly increased the rate of diagnostic failure.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Biopsy, Needle / methods*
  • Diagnostic Errors
  • Female
  • Fluoroscopy / methods*
  • Humans
  • Lung Diseases / diagnosis*
  • Lung Diseases / pathology*
  • Lung Neoplasms / diagnosis*
  • Lung Neoplasms / pathology*
  • Male
  • Middle Aged
  • Outcome Assessment, Health Care
  • Regression Analysis
  • Retrospective Studies
  • Risk Factors
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*