Thermal desorption gas chromatography-mass spectrometry as an enhanced method for the quantification of polycyclic aromatic hydrocarbons from ambient air particulate matter

J Chromatogr A. 2009 May 1;1216(18):4030-9. doi: 10.1016/j.chroma.2009.02.043. Epub 2009 Feb 21.

Abstract

Polycyclic aromatic hydrocarbons (PAHs) from ambient air particulate matter (PM) were analysed by a two-step thermal desorption (TD) injection system integrated to a gas chromatograph-mass spectrometer (GC/MS). The operational variables of the TD method were optimised and the analytical expanded uncertainties were calculated to vary from 8% to 16% over the operative concentration range (40-4000 pg). The performance of the TD method was validated by the analysis of a standard reference material and by comparison of PAH concentrations in PM samples to those obtained by a conventional liquid extraction (LE) method. The TD method reported lower uncertainties than the LE method for the analysis of similar concentrations in air. The TD method also showed advantages for shorter sampling times in comparison to 24 h for source apportionment applications and for reducing losses of more reactive compounds such as benzo[a]pyrene.

Publication types

  • Evaluation Study

MeSH terms

  • Air Pollutants / chemistry
  • Differential Thermal Analysis / methods*
  • Gas Chromatography-Mass Spectrometry / methods*
  • Particulate Matter / chemistry*
  • Polycyclic Aromatic Hydrocarbons / chemistry*

Substances

  • Air Pollutants
  • Particulate Matter
  • Polycyclic Aromatic Hydrocarbons