On the origin of the MR image phase contrast: an in vivo MR microscopy study of the rat brain at 14.1 T

Neuroimage. 2009 Jun;46(2):345-52. doi: 10.1016/j.neuroimage.2009.02.023. Epub 2009 Feb 27.

Abstract

Recent studies at high magnetic fields using the phase of gradient-echo MR images have shown the ability to unveil cortical substructure in the human brain. To investigate the contrast mechanisms in phase imaging, this study extends, for the first time, phase imaging to the rodent brain. Using a 14.1 T horizontal bore animal MRI scanner for in vivo micro-imaging, images with an in-plane resolution of 33 microm were acquired. Phase images revealed, often more clearly than the corresponding magnitude images, hippocampal fields, cortical layers (e.g. layer 4), cerebellar layers (molecular and granule cell layers) and small white matter structures present in the striatum and septal nucleus. The contrast of the phase images depended in part on the orientation of anatomical structures relative to the magnetic field, consistent with bulk susceptibility variations between tissues. This was found not only for vessels, but also for white matter structures, such as the anterior commissure, and cortical layers in the cerebellum. Such susceptibility changes could result from variable blood volume. However, when the deoxyhemoglobin content was reduced by increasing cerebral blood flow (CBF) with a carbogen breathing challenge, contrast between white and gray matter and cortical layers was not affected, suggesting that tissue cerebral blood volume (and therefore deoxyhemoglobin) is not a major source of the tissue phase contrast. We conclude that phase variations in gradient-echo images are likely due to susceptibility shifts of non-vascular origin.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain / cytology*
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / veterinary*
  • Microscopy, Phase-Contrast / methods*
  • Microscopy, Phase-Contrast / veterinary*
  • Rats
  • Rats, Sprague-Dawley
  • Reproducibility of Results
  • Sensitivity and Specificity