Linearity calibration of amplitude and power measurements in terahertz systems and detectors

Opt Lett. 2009 Mar 1;34(5):674-6. doi: 10.1364/ol.34.000674.

Abstract

We describe a device consisting of a set of silicon plates to be employed for calibrating the linearity of amplitude/power measurements of terahertz systems and detectors. The device is used to test a time-domain spectrometer, a pyroelectric sensor, and a Golay cell. The results demonstrate that the linearity of amplitude/power measurement cannot be assumed but must be experimentally verified.