4D-Imaging of the lung: reproducibility of lesion size and displacement on helical CT, MRI, and cone beam CT in a ventilated ex vivo system

Int J Radiat Oncol Biol Phys. 2009 Mar 1;73(3):919-26. doi: 10.1016/j.ijrobp.2008.09.014.

Abstract

Purpose: Four-dimensional (4D) imaging is a key to motion-adapted radiotherapy of lung tumors. We evaluated in a ventilated ex vivo system how size and displacement of artificial pulmonary nodules are reproduced with helical 4D-CT, 4D-MRI, and linac-integrated cone beam CT (CBCT).

Methods and materials: Four porcine lungs with 18 agarose nodules (mean diameters 1.3-1.9 cm), were ventilated inside a chest phantom at 8/min and subject to 4D-CT (collimation 24 x 1.2 mm, pitch 0.1, slice/increment 24 x 10(2)/1.5/0.8 mm, pitch 0.1, temporal resolution 0.5 s), 4D-MRI (echo-shared dynamic three-dimensional-flash; repetition/echo time 2.13/0.72 ms, voxel size 2.7 x 2.7 x 4.0 mm, temporal resolution 1.4 s) and linac-integrated 4D-CBCT (720 projections, 3-min rotation, temporal resolution approximately 1 s). Static CT without respiration served as control. Three observers recorded lesion size (RECIST-diameters x/y/z) and axial displacement. Interobserver- and interphase-variation coefficients (IO/IP VC) of measurements indicated reproducibility.

Results: Mean x/y/z lesion diameters in cm were equal on static and dynamic CT (1.88/1.87; 1.30/1.39; 1.71/1.73; p > 0.05), but appeared larger on MRI and CBCT (2.06/1.95 [p < 0.05 vs. CT]; 1.47/1.28 [MRI vs. CT/CBCT p < 0.05]; 1.86/1.83 [CT vs. CBCT p < 0.05]). Interobserver-VC for lesion sizes were 2.54-4.47% (CT), 2.29-4.48% (4D-CT); 5.44-6.22% (MRI) and 4.86-6.97% (CBCT). Interphase-VC for lesion sizes ranged from 2.28% (4D-CT) to 10.0% (CBCT). Mean displacement in cm decreased from static CT (1.65) to 4D-CT (1.40), CBCT (1.23) and MRI (1.16).

Conclusions: Lesion sizes are exactly reproduced with 4D-CT but overestimated on 4D-MRI and CBCT with a larger variability due to limited temporal and spatial resolution. All 4D-modalities underestimate lesion displacement.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cone-Beam Computed Tomography / methods
  • Equipment Design
  • Lung Neoplasms* / diagnostic imaging
  • Lung Neoplasms* / pathology
  • Lung* / diagnostic imaging
  • Lung* / pathology
  • Magnetic Resonance Imaging / methods
  • Movement*
  • Multiple Pulmonary Nodules* / diagnostic imaging
  • Multiple Pulmonary Nodules* / pathology
  • Observer Variation
  • Particle Accelerators
  • Phantoms, Imaging
  • Reproducibility of Results
  • Respiration*
  • Swine
  • Tomography, Spiral Computed / methods