Stem fracture of a collarless, polished, double-taper cemented femoral prosthesis: a case report

Acta Orthop Belg. 2008 Oct;74(5):697-9.

Abstract

We report a case of stem fracture of a collarless, polished, double-taper cemented stainless steel femoral prosthesis. Scanning electron microscopy showed fatigue striations, indicating failure by fatigue. This case study highlights that stem fracture can still occur with modern implants. Size 0 CPT stem should be avoided in patients with high body mass index. To our knowledge this is the first case report of a CPT stem fracture used for primary total hip replacement.

Publication types

  • Case Reports

MeSH terms

  • Arthroplasty, Replacement, Hip / instrumentation
  • Female
  • Femur
  • Hip Prosthesis*
  • Humans
  • Middle Aged
  • Prosthesis Failure