Abstract

Background and study aims: The aim of this prospective study was to compare fine-needle aspiration guided by endoscopic ultrasonography (EUS-FNA) using 25-gauge and 22-gauge needles with the EUS-guided 19-gauge Trucut needle biopsy (EUS-TNB) in patients with solid pancreatic mass.

Patients and methods: Twenty-four consecutive patients with pancreatic mass underwent biopsies by both EUS-FNA and EUS-TNB. Three needles were compared with respect to technical success rate, tissue size obtained, overall diagnostic accuracy and accuracy for histological and cytological diagnosis.

Results: The 25-gauge EUS-FNA was technically easier and obtained superior overall diagnostic accuracy than the 22-gauge and Trucut needles, especially in lesions of the pancreas head and uncinate process. Overall accuracy for the 25-gauge, 22-gauge and Trucut needle was 91.7%, 79.7% and 54.1%, respectively. Accuracy for cytological diagnosis irrespective the site of lesions with 25-gauge, 22-gauge and Trucut needles was 91.7%, 75.0%, and 45.8%, respectively. For uncinate masses, it was 100%, 33.3%, and 0.0%, respectively. These differences were significant. Among technically successful patients, the accuracy for histological diagnosis using the 25-gauge was significantly inferior (P < 0.05) to 22-gauge and Trucut needles and the rates were 45.8%, 78.9% and 83.3%.

Conclusions: The 25-gauge FNA needle was significantly superior in terms of technical success rate and overall diagnostic accuracy, especially for the head and uncinate lesions, compared to the 22-gauge and Trucut needles and could be considered 'the best choice needle for cytological diagnosis' of solid pancreatic lesions. If histological diagnosis is required, the 22-gauge FNA needle and Trucut needle may be advantageous for use in head/uncinate and body/tail lesions, respectively.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biopsy, Fine-Needle / adverse effects
  • Biopsy, Fine-Needle / instrumentation*
  • Equipment Design
  • Granuloma, Plasma Cell / diagnostic imaging
  • Granuloma, Plasma Cell / pathology*
  • Humans
  • Needles*
  • Pancreas / diagnostic imaging
  • Pancreas / pathology*
  • Pancreatic Diseases / diagnostic imaging
  • Pancreatic Diseases / pathology*
  • Pancreatic Neoplasms / diagnostic imaging
  • Pancreatic Neoplasms / pathology*
  • Predictive Value of Tests
  • Prospective Studies
  • Sensitivity and Specificity
  • Ultrasonography, Interventional*