Isolation of tetracyclines in milk using a solid-phase extracting column and water eluent

Talanta. 2003 Jan 2;59(1):155-9. doi: 10.1016/s0039-9140(02)00472-1.

Abstract

An isolating method using a solid-phase extraction (SPE) ISOLUTE(R) C8 endcapped syringe-column for routine monitoring of residual tetracyclines (TCs) (oxytetracycline (OTC), tetracycline (TC), chlortetracycline (CTC), and doxycycline (DC)) in cow's milk is presented. In the simplest and most environmentally harmless method, milk samples could be applied directly to the SPE column, following which all TCs were eluted with water. No organic solvents were used at all. The purified sample was injected into a high-performance liquid chromatography (HPLC) with a photo-diode array detector (PDAD). For the HPLC determination/identification, a LiChrospher(R) 100 RP-8 endcapped column and a mobile phase of acetonitrile -7% (v v(-1)) acetic acid solution (in water) (35:65, v v(-1)) with a PDAD was used. The total time required for the analysis of one sample was <40 min. Average recoveries (spiked 0.1-1.0 mug ml(-1) each drug) and their standard deviations were >80 and <5%, respectively.