Method validation for determination of arsenic, cadmium, chromium and lead in milk by means of dynamic reaction cell inductively coupled plasma mass spectrometry

Anal Chim Acta. 2008 Aug 22;624(1):59-67. doi: 10.1016/j.aca.2008.06.024. Epub 2008 Jun 24.

Abstract

With Regulation No. 1881/2006 the European Union fixed a maximum level for lead in milk. Consequently, there is the need to determine very low concentration of elements that may be present in milk in trace and ultratrace levels. Quadrupole inductively coupled plasma mass spectrometry (Q-ICP-MS) combined with dynamic reaction cell (DRC) has been widely employed in order to reach very low concentration, requested for this product. Furthermore, the DRC technology can help in removing polyatomic and argon-based interferences. In the present study, a method for the determination of arsenic, cadmium, chromium and lead in bovine milk was validated according to the EU common standards by means of DRC-ICP-MS. The main parameters evaluated in the validation were: recovery, repeatability and within-laboratory reproducibility, detection and quantification limits, linearity range and measurement uncertainty. Additionally, stability studies of the analyte in solution and ruggedness studies were carried out. The results obtained for limit of detection (LoD) and limit of quantification (LoQ) in microg kg(-1) were respectively: As, 3.1 and 9.5; Cd, 0.08 and 0.24; Cr, 0.229 and 0.693; Pb, 0.5 and 1.5. While for the recovery: As, 91%; Cd 96%; Cr 99%; Pb, 95%. As for the repeatability: As, 7%; Cd, 3%; Cr, 6%; Pb, 4%.

Publication types

  • Validation Study

MeSH terms

  • Animals
  • Arsenic / analysis*
  • Cadmium / analysis*
  • Calibration
  • Chromium / analysis*
  • European Union
  • Food Contamination / analysis*
  • Humans
  • Infant Formula / chemistry
  • Lead / analysis*
  • Mass Spectrometry / methods
  • Milk / chemistry*
  • Milk, Human / chemistry
  • Reproducibility of Results

Substances

  • Cadmium
  • Chromium
  • Lead
  • Arsenic