The comparison of doses measured by radiochromic films and semiconductor detector in a 175MeV proton beam

Phys Med. 2009 Sep;25(3):105-10. doi: 10.1016/j.ejmp.2008.06.001. Epub 2008 Aug 10.

Abstract

We wanted to verify the response of radiochromic films in a 175MeV clinical proton beam used at the Joint Institute for Nuclear Research in Dubna against doses measured using semiconductor detectors and compare the results with published data from other centres. Radiochromic films (RCFs) MD-55 and a Vidar VXR-16 scanner were used. The films were irradiated in an unmodulated proton beam and with a beam modulated with a bolus and a ridge filter. Obtained dose distributions were compared with dose distributions measured with a Si-semiconductor detector. For the unmodulated beam the difference between the RCF and the semiconductor detector was 12% in the Bragg peak top. For the modulated beam the difference inside the spread-out Bragg peak region was 4%. Observed deviations between doses measured with RCF and Si-detector outside the Bragg peak were caused by the inhomogeneity of radiochromic emulsion. In the Bragg peak region the RCF doses were lower than those measured by semiconductors. The results were in agreement with published data from other proton therapy centres.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Film Dosimetry / instrumentation*
  • Proton Therapy*
  • Reproducibility of Results
  • Semiconductors*
  • Sensitivity and Specificity
  • Transducers*