Detection of significant coronary artery stenosis with 64-slice computed tomography in heart transplant recipients: a comparative study with conventional coronary angiography

Int J Cardiovasc Imaging. 2009 Jan;25(1):91-100. doi: 10.1007/s10554-008-9343-z. Epub 2008 Jul 19.

Abstract

Purpose: The present study evaluates clinical feasibility of cardiac multidetector computed tomography angiography (MDCTA) to detect significant stenosis of coronary vessels due to transplant vasculopathy (TVP) after heart transplantation (HTx).

Methods: Twenty-eight consecutive male HTx-recipients scheduled for their annual routine conventional coronary angiography (CCA) additionally underwent 64-slice MDCTA.

Results: Two patients were excluded from further MDCTA analysis. Out of 371 remaining coronary vessel segments evaluable by CCA, MDCTA was able to depict 302 (81.4%) in diagnostic image quality. On a segment based analysis, sensitivity, specificity, diagnostic accuracy (DA), negative predictive value (NPV), and positive predictive value (PPV) for detection of significant stenosis were calculated with 87.5%, 97.3%, 97.0%, 99.7%, and 46.7%, respectively. On a patient-based evaluation, sensitivity, specificity, DA, NPV, PPV were 100%, 81%, 84.6%, 100% and 55.6%, respectively. Evaluation of stenosis degree by MDCTA showed systematic overestimation of 4.4%. A moderate to good agreement comparing both modalities was found (Pearson's correlation coefficient: 0.64).

Conclusion: High NPV suggesting 64-slice MDCTA being a reliable diagnostic tool for ruling out significant stenosis due to TVP in HTx patients. But its clinical value in these particular patients needs further investigation.

Publication types

  • Comparative Study

MeSH terms

  • Coronary Angiography / methods*
  • Coronary Stenosis / diagnostic imaging*
  • Heart Transplantation*
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications / diagnostic imaging*
  • Predictive Value of Tests
  • Radiographic Image Interpretation, Computer-Assisted
  • Regression Analysis
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*