Alternative antibody for the detection of CA15-3 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access BR Monitor assay on the UniCel Dxl 800 Immunoassay System

Clin Chem Lab Med. 2008;46(5):612-22. doi: 10.1515/CCLM.2008.133.

Abstract

Background: Cancer antigen CA15-3 antigen is known as a valuable marker for the management of breast cancer.

Methods: The analytical and clinical performance of the Access BR Monitor Immunoassay System (Beckman Coulter) was evaluated at five different European sites and compared with a reference system, defined as CA15-3 on the Elecsys System (Roche Diagnostics).

Results: Total imprecision (% CV) of the BR Monitor ranged between 5.5% and 11.7%, and inter-laboratory reproducibility between 3.4% and 5.1%. Linearity upon dilution showed a mean recovery of 98.5% (SD + 9.1%). Endogenous interferents had no influence on BR Monitor levels (mean recoveries: hemoglobin 112%, bilirubin 111%, triglycerides 108%). There was no high-dose hook effect up to 13,540 kU/L. Clinical performance investigated in sera from individuals showed a general correlation between the Access BR Monitor and Elecsys CA15-3 (R = 0.797), with a slope of 1.383. CA15-3 serum levels, as measured by the BR Monitor, were low in healthy individuals (n = 267, median = 11.9 kU/L, 95th percentile = 23.5 kU/L), higher in individuals with various benign diseases (n = 549, medians = 11.3-15.6 kU/L, 95th percentiles = 21.6-54.6 kU/L) and even higher in individuals suffering from various cancers (n = 995, medians = 11.2-22.8 kU/L, 95th percentiles = 30.0-429.7 kU/L). Best diagnostic accuracy for cancer detection against the relevant benign control group by the BR Monitor was found for locoregional and metastatic breast cancer, as well as for ovarian cancer [area under the curve (AUC) 0.619, 0.897 and 0.774]. Results for the reference CA15-3 assay were comparable (AUC 0.611, 0.887 and 0.818).

Conclusions: The Access BR Monitor provides accurate methodological characteristics and demonstrates an analytical and clinical correlation with Elecsys CA15-3. Best diagnostic accuracy for the BR Monitor was found in breast and ovarian cancer. Our results also suggest a clinical value of the BR Monitor in other cancers.

Publication types

  • Comparative Study
  • Evaluation Study
  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Antibodies / chemistry*
  • Area Under Curve
  • Breast Neoplasms / blood
  • Breast Neoplasms / diagnosis*
  • Europe
  • Female
  • Humans
  • Immunoassay / instrumentation
  • Immunoassay / methods*
  • Middle Aged
  • Mucin-1 / blood*
  • Ovarian Neoplasms / blood
  • Ovarian Neoplasms / diagnosis*
  • Reference Standards
  • Reproducibility of Results