Performance of the GeneXpert enterovirus assay for detection of enteroviral RNA in cerebrospinal fluid

J Clin Virol. 2008 Sep;43(1):110-3. doi: 10.1016/j.jcv.2008.04.006. Epub 2008 Jun 4.

Abstract

Background: The GeneXpert Dx System allows for automated extraction, processing, amplification and real-time detection of target nucleic acids.

Objectives: To evaluate the performance of the Cepheid Xpert enterovirus (EV) assay for detection of EV RNA compared to a nucleic acid sequence based amplification (NASBA) assay and a user-developed TaqMan RT-PCR assay.

Study design: Assays were evaluated using a 12-member proficiency panel and up to 138 CSF specimens. Samples in which EV RNA was detected by two or more assays were considered true positives.

Results: The GeneXpert, NASBA, and TaqMan assays correctly identified 10, 8, and 7 of 12 proficiency panel members, respectively. For detection of EV RNA in CSF, the sensitivities of the GeneXpert, NASBA, and TaqMan were 100%, 87.5%, and 96%, respectively. There were no false positives. Two samples tested by GeneXpert and NASBA yielded indeterminate or invalid results and could not be resolved.

Conclusions: The Xpert EV assay is a sensitive and specific method for detection of EV RNA in CSF specimens. The ease of use, random access capability, and minimal hands-on time with the automated GeneXpert system affords laboratories with little molecular diagnostics expertise an opportunity to complete a clinically useful testing within 2.5 h.

Publication types

  • Evaluation Study

MeSH terms

  • Enterovirus / genetics
  • Enterovirus / isolation & purification*
  • Enterovirus Infections / cerebrospinal fluid
  • Enterovirus Infections / diagnosis*
  • Humans
  • Molecular Diagnostic Techniques / methods*
  • RNA, Viral / cerebrospinal fluid*
  • Reverse Transcriptase Polymerase Chain Reaction / methods
  • Self-Sustained Sequence Replication / methods
  • Sensitivity and Specificity

Substances

  • RNA, Viral