Analysis of polybrominated diphenyl ethers (PBDEs) by liquid chromatography with negative-ion atmospheric pressure photoionization tandem mass spectrometry (LC/NI-APPI/MS/MS): application to house dust

Anal Bioanal Chem. 2008 Jul;391(6):2249-56. doi: 10.1007/s00216-008-2156-z. Epub 2008 May 27.

Abstract

Eight polybrominated diphenyl ether (PBDE) congeners of primary interest to the US EPA were separated using reverse-phase liquid chromatography on an octadecylsilane column. BDE-28, BDE-47, BDE-99, BDE-100, BDE-153, BDE-154, BDE-183, and BDE-209 were baseline-resolved under isocratic conditions in 92:8 methanol/water (v/v). Negative-ion atmospheric pressure photoionization (NI-APPI) with a toluene dopant produced precursor ions corresponding to [M-Br+O](-) for the eight congeners studied. Each congener was quantified by tandem mass spectrometry through a unique multiple reaction monitoring (MRM) transition. On-column limits of detection were between 2.4 and 27.8 pg for the eight congeners studied, with an intra-day method precision of 9%. The LC/NI-APPI/MS/MS method was validated for the analysis of the eight PBDE congeners in NIST SRM 2585 (Organics in House Dust). Pressurized liquid extraction (PLE) with subsequent LC/NI-APPI/MS/MS analysis afforded quantitative recovery for all eight PBDE congeners with recoveries ranging from 92.7 to 113%. The liquid-phase separation of the LC/NI-APPI/MS/MS method is not prone to the thermal degradation issues that plague splitless GC based analyses of highly brominated PBDEs such as BDE-209.