Abstract

Objective: To evaluate the radiation field feature of Re esophageal stent and provide scientific basis for clinical application.

Methods: We measure the beta-ray, gamma-ray and bremssfrahlung dose of every selected point on the bionics esophageal stent and then draw out computer software by mathematical formula.

Results: The radiation field of Re esophageal stent has its own feature: the max range of beta-ray is 11 mm, 90% dose construction field is within 1.5 mm, 95% dose range is within 2.5 mm, and only 4.21% of total energy of gamma-ray and bremssfrahlung is out of 6.5 mm range. The absorption dose of every direction in same point of the esophageal model was similar (P>0.05).

Conclusion: Beta ray is the major radiation of Re esophageal stent while gamma-ray and bremssfrahlung are 4.21% among the radiation field. The max dose construction field is within 0.5-1.5 mm, just short at the depth of esophagus mucosa within 0.5-1.5 mm range. So Re stent is a good choice of palliative intracavitary radiotherapy of esophageal carcinoma.

Publication types

  • Evaluation Study

MeSH terms

  • Brachytherapy / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Esophagus / surgery*
  • Radiometry / methods*
  • Radiotherapy Dosage*
  • Stents*