Development of an amperometric flow analysis sensor for specific detection of D-psicose

Biosens Bioelectron. 2008 Apr 15;23(9):1347-52. doi: 10.1016/j.bios.2007.12.003. Epub 2007 Dec 14.

Abstract

The aim of this study was to develop a simple, rapid and highly sensitive sensor for measuring the rare sugar d-psicose. The proposed system adopts amperometric flow analysis and two consecutive enzyme reactions consisting of a reactor packed with d-tagatose 3-epimerase (DTE)-immobilized beads, which converts d-psicose to d-fructose, and a carbon-paste electrode containing d-fructose dehydrogenase (DFDH). In order to fabricate a robust sensor system, various experimental parameters were optimized including the buffer composition, flow rate for the two enzyme reactions and the size of micro-flow cell. The developed sensor responded linearly to d-psicose concentration in the range from 0.08 to 50mM (R(2)=0.988). The signal/noise ratio was 3.0 for the 0.08 mM d-psicose solution, and the relative standard deviations were 1.7 (n=20) and 2.6% (n=20) for the 10 and 20mM d-psicose solutions, respectively. One round of assay was completed within 8 min. Our results suggest that the sensor can be used not only for the detection of d-psicose in food samples but also for monitoring d-psicose within the environment. Moreover, the sensor system can be applied to the detection of many other rare sugars by using the same measurement principle.

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Biosensing Techniques / methods
  • Electrochemistry
  • Flow Injection Analysis / instrumentation*
  • Fructose / analysis*
  • Reproducibility of Results

Substances

  • psicose
  • Fructose