Capacitive micromachined ultrasonic transducers (CMUTs) with isolation posts

Ultrasonics. 2008 Mar;48(1):74-81. doi: 10.1016/j.ultras.2007.11.006. Epub 2007 Dec 14.

Abstract

In this paper, an improved design of a capacitive micromachined ultrasonic transducer (CMUT) is presented. The design improvement aims to address the reliability issues of a CMUT and to extend the device operation beyond the contact (collapse) voltage. The major design novelty is the isolation posts in the vacuum cavities of the CMUT cells instead of full-coverage insulation layers in conventional CMUTs. This eliminates the contact voltage drifting due to charging caused by the insulation layer, and enables repeatable CMUT operation in the post-contact regime. Ultrasonic tests of the CMUTs with isolation posts (PostCMUTs) in air (electrical input impedance and capacitance vs. bias voltage) and immersion (transmission and reception) indicate acoustic performance similar to that obtained from conventional CMUTs while no undesired side effects of this new design is observed.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Calibration
  • Computer Simulation
  • Computer-Aided Design*
  • Electric Capacitance
  • Electronics, Medical
  • Equipment Design
  • Equipment Failure Analysis
  • Microelectrodes*
  • Miniaturization
  • Transducers*
  • Ultrasonography / instrumentation*